55
Thorax
2
nd
reconstr.
kV
110
Effective mAs
*
Slice collimation
1.0 mm
Slice width
3.0 mm
1.25 mm
Feed/Rotation
9.0 mm
Rotation time
0.8 sec.
Kernel
B41s
B60s
Increment
3.0 mm
0.8 mm
Image order
cr-ca
Tips
• Children with a body weight of more than 55 kg
should be examined with an adult protocol.
• Two recon jobs are predefined for image reconstruc-
tion of routine thorax and thin slice lung.
Contrast medium IV injection
Start delay
exam dependent
Flow rate
dependent upon needle size/Access site
Total amount
1 – 2 ml per kg of body weight
* The mAs should be adjusted to the body weight
of the child.
Body weight
kV
mAs
CTDI
Vol
(mGy)
< 15 kg
110
20
15 – 24 kg
110
26
2.4
25 – 34 kg
110
39
3.5
35 – 44 kg
110
65
45 – 55 kg
110
90
8.2
Children