Do not administer oral contrast medium, as this
impairs the editing of MIP/SSD/VRT images. Use water
instead if necessary.
* CARE Bolus may be used to optimize the bolus timing.
Set the ROI for monitoring scan in the abdominal
aorta at the beginning of the scan range with trigger-
ing threshold of 120 HU, or use manual triggering.
MIPthin: ArterialPhase
coronal
Image thickness
3
Distance between images
3
Number of images
50
For the 2
nd
reconstruction of the Arterial Phase the
Autoload into the MIPthin Range on the 3D Card is acti-
vated. The images will be automatically loaded into
3D, MIPthin, and coronal MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
Contrast medium IV injection
Start delay
20 – 25 sec.* (arterial phase)
50 – 75 sec.* (venous phase)
Flow rate
4 – 5 ml/sec.
Total amount
100 – 120 ml
149
Abdomen
Non
Arterial
Venous
Contrast
phase
phase
kV
110
110
130
Effective mAs
60
90
95
Slice collimation 2.0 mm
2.0 mm
2.0 mm
Slice width
5.0 mm
5.0 mm/
5.0 mm
2.5 mm
Feed/Rotation
18.0 mm
12.0 mm
12.0 mm
Rotation time
0.8 sec.
0.8 sec.
0.8 sec.
Kernel
B41s
B41s/B31s B41s
Increment
5.0 mm
5.0 mm/
5.0 mm
1.5 mm
Image order
cr-ca
cr-ca
cr-ca
CTDI
Vol
4.7 mGy
7.0 mGy
10.8 mGy
Summary of Contents for SOMATOM Emotion 6
Page 1: ...SOMATOM Emotion 6 Application Guide Routine Protocols Software Version A70 ...
Page 23: ...23 General ...
Page 43: ...43 Children ...
Page 59: ...59 Children ...
Page 89: ...Head 89 ...
Page 102: ...102 Shoulder Overview Shoulder Spiral mode for HiRes bone studies and soft tissues ...
Page 139: ...139 Abdomen ...
Page 167: ...167 Pelvis ...
Page 175: ...175 Spine ...
Page 185: ...185 Extremities ...