97
Neck
AngioCarotid
2
nd
reconstr.
kV
110
Effective mAs
70
Slice collimation
1.0 mm
Slice width
3.0 mm
1.25 mm
Feed/Rotation
9.0 mm
Rotation time
0.6 sec.
Kernel
B31s
B31s
Increment
3.0 mm
0.8 mm
Image order
cr-ca
CTDI
Vol
6.4 mGy
CARE Bolus may be used to optimize the bolus timing.
Set the ROI for monitoring scan in the aortic arch
with triggering threshold of 120 HU, or use manual
triggering.
MIPthin: AngioCarotid
coronal
Image thickness
3
Distance between images
3
Number of images
20
For the 2
nd
reconstruction the Autoload into MIPthin
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MIPthin, and a coronal
MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
Contrast medium IV injection
Start delay
15 sec.
Flow rate
4 ml/sec.
Total amount
90 ml