
179
Spine
Spine
2
nd
reconstr.
kV
130
Effective mAs
190
Slice collimation
2.0 mm
Slice width
2.5 mm
2.5 mm
Feed/Rotation
8.0 mm
Rotation time
1.0 sec.
Kernel
B31s
B60s
Increment
2.5 mm
1.5 mm
Image order
cr-ca
CTDI
Vol
21.6 mGy
MPRthick: Spine
coronal
Image thickness
3
Distance between images
3
Number of images
30
For the 2
nd
reconstruction the Autoload into the MPR
Range on the 3D Card is activated. The images will
be automatically loaded into 3D, MPR, and a coronal
MPRthick Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
Summary of Contents for SOMATOM Emotion 6
Page 1: ...SOMATOM Emotion 6 Application Guide Routine Protocols Software Version A70 ...
Page 23: ...23 General ...
Page 43: ...43 Children ...
Page 59: ...59 Children ...
Page 89: ...Head 89 ...
Page 102: ...102 Shoulder Overview Shoulder Spiral mode for HiRes bone studies and soft tissues ...
Page 139: ...139 Abdomen ...
Page 167: ...167 Pelvis ...
Page 175: ...175 Spine ...
Page 185: ...185 Extremities ...