Appendi x F Acoustic Output Reference
F - 4
Instructions for Use
Default Displayed MI and TI Values by Transducer
(Per transducer/mode that exceeds default MI or TI value of 0.4)
Refer to the Chapter 1 and Appendix A in this manual to identify transducers compatible with
your ultrasound system.
General Imaging
Mode
B
M
PWD
SCW
Color
CWD
Transducer
MI
TI
MI
TI
MI
TI
MI
TI
MI
TI
MI
TI
4C1
1.2
-
1.4
2.2(B
2
) 0.8
3.4(B
2
)
-
-
1.1
0.5(S
1
) -
-
6C1 HD
1.4
0.5(S
1
) 1.4
1.6(B
2
) 1.1
4.0(B
2
)
-
-
1.4
0.8(S
1
) -
-
6C2
1.0
0.5(S
1
) 1.4
1.2(B
2
) 0.7
3.1(B
2
)
-
-
1.0
0.8(S
1
) -
-
8C3 HD
0.6
-
0.7
-
0.7
1.4(B
2
)
0.4(S
1
)
-
-
0.8
-
-
-
9L4
1.0
-
1.0
-
1.4
1.3(B
2
)
-
-
0.9
1.9(S
1
) -
-
12L4
0.8
-
0.8
-
1.3
0.6(B2)
0.6(S1)
-
-
0.8
0.6(B2)
0.6(S1)
-
-
14L5
1.0
-
0.9
-
1.1
1.8(B
2
)
-
-
0.9
1.2(S
1
) -
-
14L5 SP
0.8
-
0.9
-
0.6
1.0(B
2
)
-
-
1.3
0.8(S
1
) -
-
18L6 HD
1.0
-
0.7
-
1.0
1.2(B
2
)
-
-
1.0
0.6(S
1
) -
-
4P1
1.2
1.0(S
1
) 1.4
1.9(B
2
) 1.0
3.8(B
2
)
-
-
1.2
1.2(S
1
) -
-
4V1
0.8
0.4(S
1
) 1.1
1.4(B
2
) 0.9
4.3(B
2
)
-
-
0.8
0.8(S
1
) -
-
4V1c
1.4
0.9(S
1
) 1.4
1.4(B
2
) 1.0
3.7(B
2
)
-
-
1.4
1.1(S
1
) -
-
8V3
0.6
0.5(S
1
) 1.2
0.4(B
2
) 1.4
2.5(B
2
)
-
-
0.7
0.9(S
1
) -
-
10V4
0.8
1.0(S
1
) 1.5
0.4(B
2
) 0.6
1.9(B
2
)
-
-
0.8
1.2(S
1
) -
-
EV-8C4
0.9
0.5(S
1
) 1.3
0.7(B
2
) 1.2
2.0(B
2
)
-
-
1.0
0.6(S
1
) -
-
EC9-4
0.4
-
0.4
-
0.6
1.0(B
2
)
-
-
0.8
0.4(S
1
) -
-
MC9-4
0.8
-
0.8
-
1.0
1.8(B
2
)
0.9(S
1
)
-
-
0.7
-
-
-
7CF1
1.0
-
1.2
0.9(B
2
)
0.5(S
1
)
0.5
1.7(B
2
)
0.8(S
1
)
-
-
0.9
0.5(B
2
)
0.5(S
1
)
-
-
7CF2
1.1
-
1.1
0.8(B
2
) 0.5
1.8(B
2
)
-
-
1.2
0.5(S
1
) -
-
9EVF4
0.5
-
0.5
-
0.9
1.4(B
2
)
-
-
0.8
-
-
-
CW2
-
-
-
-
-
-
-
-
-
-
0.1 4.9(B
2
)
CW5
-
-
-
-
-
-
-
-
-
-
0.1 3.0(B
2
)
S
1
: Soft Tissue Thermal Index, B
2
: Bone Thermal Index, C
3
: Cranial Thermal Index
Summary of Contents for Acuson S2000
Page 12: ...1 Introduction 1 2 Instructions for Use ...
Page 14: ...1 Introduction 1 4 Instructions for Use System Review Example of the ultrasound system ...
Page 84: ...2 Safety and Care 2 54 Instructions for Use ...
Page 86: ...3 System Setup 3 2 Instructions for Use ...
Page 112: ...3 System Setup 3 28 Instructions for Use ...
Page 114: ...4 Examination Fundamentals 4 2 Instructions for Use ...
Page 144: ...5 Transducer Accessories and Biopsy 5 2 Instructions for Use ...
Page 196: ...7 Specialty Transducers 7 2 Instructions for Use ...
Page 200: ...7 Specialty Transducers 7 6 Instructions for Use ...
Page 202: ...8 Physiologic Function 8 2 Instructions for Use ...
Page 208: ...9 eSieFusion Imaging 9 2 Instructions for Use ...
Page 236: ...10 Virtual Touch Applications 10 2 Instructions for Use ...
Page 258: ...10 Virtual Touch Applications 10 24 Instructions for Use ...
Page 302: ...Appendix A Technical Description A 44 Instructions for Use ...
Page 326: ...Appendix B Control Panel and Touch Screen B 24 Instructions for Use ...
Page 328: ...Appendix C Control Panel C 2 Instructions for Use ...
Page 394: ...Appendix D On screen Controls D 50 Instructions for Use ...
Page 444: ...Appendix F Acoustic Output Reference F 2 Instructions for Use ...
Page 516: ...Appendix F Acoustic Output Reference F 74 Instructions for Use ...
Page 517: ......
Page 518: ......