9 eSieFusi on Im aging
Instructions for Use
9 - 13
To
Do this
Apply predefined optimization
settings to the displayed
reference data
(Available only during planning)
Note:
If the volume quadrant is selected, the system applies the setting to the
volume data. If an MPR quadrant is selected, the system applies the setting to all
MPRs.
Select the optimization setting.
●
Rotate [
Preset
].
○
For systems without a touch screen, select a
Presets
setting on the image
menu.
Note:
CTA
is a computed tomography angiogram.
T1
and
T2
are types of
magnetic resonance imaging (MRI) scans.
Adjust the range of values
displayed in the reference data
(Available only during planning)
Note:
If the volume quadrant is selected, the system applies the setting to the
volume data. If an MPR quadrant is selected, the system applies the setting to all
MPRs.
Select the value range setting.
●
Drag the
Window and Level
icon on the touch screen to adjust the setting.
○
For systems without a touch screen, select a
Window
setting on the image
menu.
Adjust the midpoint of values
displayed in the reference data
(Available only during planning)
Note:
If the volume quadrant is selected, the system applies the setting to the
volume data. If an MPR quadrant is selected, the system applies the setting to all
MPRs.
Select the value setting.
●
Drag the
Window and Level
icon on the touch screen to adjust the setting.
○
For systems without a touch screen, select a
Level
setting on the image
menu.
Adjust the opacity of the
displayed reference data
(Available only when a blended image is displayed)
Select an opacity setting.
●
Rotate [
Transparency
].
○
For systems without a touch screen, select a
Transparency
setting on the
image menu.
The following transparency settings are available when a blended image is
displayed.
0% - The reference data is not visible.
Less than 50% - More of the ultrasound image and less of the reference data is
visible.
50% - Both the ultrasound image and the reference data are equally visible.
More than 50% - More of the reference data and less of ultrasound image is
visible.
100% - The reference data is fully visible and the ultrasound image is less
visible.
Summary of Contents for Acuson S2000
Page 12: ...1 Introduction 1 2 Instructions for Use ...
Page 14: ...1 Introduction 1 4 Instructions for Use System Review Example of the ultrasound system ...
Page 84: ...2 Safety and Care 2 54 Instructions for Use ...
Page 86: ...3 System Setup 3 2 Instructions for Use ...
Page 112: ...3 System Setup 3 28 Instructions for Use ...
Page 114: ...4 Examination Fundamentals 4 2 Instructions for Use ...
Page 144: ...5 Transducer Accessories and Biopsy 5 2 Instructions for Use ...
Page 196: ...7 Specialty Transducers 7 2 Instructions for Use ...
Page 200: ...7 Specialty Transducers 7 6 Instructions for Use ...
Page 202: ...8 Physiologic Function 8 2 Instructions for Use ...
Page 208: ...9 eSieFusion Imaging 9 2 Instructions for Use ...
Page 236: ...10 Virtual Touch Applications 10 2 Instructions for Use ...
Page 258: ...10 Virtual Touch Applications 10 24 Instructions for Use ...
Page 302: ...Appendix A Technical Description A 44 Instructions for Use ...
Page 326: ...Appendix B Control Panel and Touch Screen B 24 Instructions for Use ...
Page 328: ...Appendix C Control Panel C 2 Instructions for Use ...
Page 394: ...Appendix D On screen Controls D 50 Instructions for Use ...
Page 444: ...Appendix F Acoustic Output Reference F 2 Instructions for Use ...
Page 516: ...Appendix F Acoustic Output Reference F 74 Instructions for Use ...
Page 517: ......
Page 518: ......