10 Virtual Touch Applic ati ons
10 - 18
Instructions for Use
To activate Virtual Touch IQ:
1. During 2D-mode imaging, select the control assigned to activating the feature:
●
Press
E
and then select
Virtual Touch IQ
.
○
For systems without a touch screen, select
Virtual Touch IQ
.
Virtual Touch IQ button.
The system activates a full-screen format and displays a region of interest (ROI).
2. To activate a dual display format:
●
Select
Dual
.
○
For systems without a touch screen, double-click
UPDATE
.
The left image displays the ROI. The right image is blank.
To
Do This
Adjust the size or position of
the region of interest (ROI)
prior to acquisition
1. Activate the positioning or sizing function:
●
Press
SELECT
.
○
For systems without a touch screen, press
NEXT
.
2. Roll the trackball to reposition or resize the ROI.
Begin acquisition
Note:
You cannot adjust the ROI during acquisition.
Press
UPDATE
.
The system acquires a single image frame and then freezes the image and applies
the velocity image display setting. During system freeze, a timer indicates the
remaining time (in seconds) until the transducer is available for image acquisition.
Change the image display
setting
Rotate [
Shear Wave
] to select the required setting:
–
Velocity
depicts the range of shear wave velocity or elasticity values from
minimum to maximum. Shear wave velocity is slower in soft tissue and faster in
stiff tissue. Elasticity is higher in soft tissue and lower in stiff tissue. The color
bar displays the minimum and maximum range of the shear wave velocity and
elasticity measurement.
–
Quality
depicts the combination of shear wave displacement magnitude and the
signal-to-noise level from low (
LO
) to high (
HI
). A low displacement magnitude
and high noise level are depicted as low quality. A high displacement magnitude
and low noise level are depicted as high quality.
–
Time
indicates the travel time of the shear wave from slow (
S
) to fast (
F
).
–
Displacement
indicates the magnitude of the shear wave displacement of
tissue from low (
LO
) to high (
HI
). Displacement is higher in soft tissue and lower
in stiff tissue.
Reactivate Virtual Touch IQ
acquisition after the
automatic freeze period
ends
1. Unfreeze the image.
2. Adjust the ROI size and position, if necessary.
3. Press
UPDATE
to begin acquisition.
Exit Virtual Touch IQ
Press
2D
.
Summary of Contents for Acuson S2000
Page 12: ...1 Introduction 1 2 Instructions for Use ...
Page 14: ...1 Introduction 1 4 Instructions for Use System Review Example of the ultrasound system ...
Page 84: ...2 Safety and Care 2 54 Instructions for Use ...
Page 86: ...3 System Setup 3 2 Instructions for Use ...
Page 112: ...3 System Setup 3 28 Instructions for Use ...
Page 114: ...4 Examination Fundamentals 4 2 Instructions for Use ...
Page 144: ...5 Transducer Accessories and Biopsy 5 2 Instructions for Use ...
Page 196: ...7 Specialty Transducers 7 2 Instructions for Use ...
Page 200: ...7 Specialty Transducers 7 6 Instructions for Use ...
Page 202: ...8 Physiologic Function 8 2 Instructions for Use ...
Page 208: ...9 eSieFusion Imaging 9 2 Instructions for Use ...
Page 236: ...10 Virtual Touch Applications 10 2 Instructions for Use ...
Page 258: ...10 Virtual Touch Applications 10 24 Instructions for Use ...
Page 302: ...Appendix A Technical Description A 44 Instructions for Use ...
Page 326: ...Appendix B Control Panel and Touch Screen B 24 Instructions for Use ...
Page 328: ...Appendix C Control Panel C 2 Instructions for Use ...
Page 394: ...Appendix D On screen Controls D 50 Instructions for Use ...
Page 444: ...Appendix F Acoustic Output Reference F 2 Instructions for Use ...
Page 516: ...Appendix F Acoustic Output Reference F 74 Instructions for Use ...
Page 517: ......
Page 518: ......