2 Safety and Care
Instructions for Use
2 - 43
Required Task
Location in This
Manual
Read and understand the biohazard considerations.
p. 2-13
Perform the steps in the daily checklist.
p. 2-24
Read and understand the cleaning and disinfecting information for the ultrasound system.
p. 2-24
Read and understand all transducer-specific warnings and cautions.
p. 2-36
Disconnect the transducer.
Chapter 3
Clean and disinfect the transducer after each use.*
p. 2-43
Select an approved disinfectant.
p. 2-45
Dry and store the transducer in its protective carrying case.
p. 2-37
*Information on cleaning and disinfecting transesophageal transducers is located in Chapter 6 in this
manual.
To clean and disinfect a transducer:
1. Wipe visible gel or particles from the transducer with a soft cloth.
2. Remove any transducer accessories from the transducer.
3. Disconnect and remove the transducer from the ultrasound system.
4. To remove existing particles from the transducer prior to disinfection, moisten a clean
gauze pad or cloth with water or an approved cleaner and then wipe the transducer. Avoid
touching the electrical components on the connector.
See also:
For a list of approved cleaners, refer to p. 2-44.
5. Select an approved disinfectant for low-level or high-level disinfection. Follow the
disinfectant manufacturer's instructions for the required level of disinfection.
See also:
For a list of approved disinfectants, refer to p. 2-45.
a. Moisten a clean gauze pad with the disinfectant, if necessary.
b. Carefully wipe the transducer from the connector to the surface which was in contact
with the patient. Avoid touching the electrical components on the connector.
Note:
If immersion is required, keep the transducer connector and connector strain relief dry
while immersing the transducer in an approved disinfectant to the immersion level indicated on
p. 2-41.
6. Follow the disinfectant manufacturer's instructions for drying the transducer.
Note:
If the disinfectant manufacturer does not provide drying instructions, dry the transducer
with a clean, lint-free, soft cloth.
Summary of Contents for Acuson S2000
Page 12: ...1 Introduction 1 2 Instructions for Use ...
Page 14: ...1 Introduction 1 4 Instructions for Use System Review Example of the ultrasound system ...
Page 84: ...2 Safety and Care 2 54 Instructions for Use ...
Page 86: ...3 System Setup 3 2 Instructions for Use ...
Page 112: ...3 System Setup 3 28 Instructions for Use ...
Page 114: ...4 Examination Fundamentals 4 2 Instructions for Use ...
Page 144: ...5 Transducer Accessories and Biopsy 5 2 Instructions for Use ...
Page 196: ...7 Specialty Transducers 7 2 Instructions for Use ...
Page 200: ...7 Specialty Transducers 7 6 Instructions for Use ...
Page 202: ...8 Physiologic Function 8 2 Instructions for Use ...
Page 208: ...9 eSieFusion Imaging 9 2 Instructions for Use ...
Page 236: ...10 Virtual Touch Applications 10 2 Instructions for Use ...
Page 258: ...10 Virtual Touch Applications 10 24 Instructions for Use ...
Page 302: ...Appendix A Technical Description A 44 Instructions for Use ...
Page 326: ...Appendix B Control Panel and Touch Screen B 24 Instructions for Use ...
Page 328: ...Appendix C Control Panel C 2 Instructions for Use ...
Page 394: ...Appendix D On screen Controls D 50 Instructions for Use ...
Page 444: ...Appendix F Acoustic Output Reference F 2 Instructions for Use ...
Page 516: ...Appendix F Acoustic Output Reference F 74 Instructions for Use ...
Page 517: ......
Page 518: ......