![Shenzhen Mindray Bio-Medical Electronics Co., Ltd. DC-N2 Operator'S Manual Download Page 48](http://html1.mh-extra.com/html/shenzhen-mindray-bio-medical-electronics-co-ltd/dc-n2/dc-n2_operators-manual_1256057048.webp)
3-10 System Preparation
h
Select the printer model from the drop-down list right to the “Printer”;
h
Set “Paper Size”.
6. Click [Save] to finish the installation.
Tips: Drivers of some popular printers have already been integrated in the system, and will be
installed automatically. You need to check the following information to check if the auto-
installation fails:
z
Model of the connected printer is not displayed in the “Printer Driver” list;
z
Click “Printer Driver”->[Add Local Printer], the system will prompts you to update the
ppd. file.
At this time, you need to install the printer driver manually as described in the following:
a) Download the ppd. file from the printer manufacturer’s official website (contact
service engineer if necessary), and copy the ppd. file to the storage device (USB
disk as an example).
b) Connect the USB disk to the ultrasound system, click [Setup]->[Print Preset]-
>”Printer Driver”->[Add Local Printer] to run the ppd. file and finish the installation.
NOTE:
All printer drivers integrated in the ultrasound system come from official website of
each printer manufacturer (you can check models in the [Setup]->[Print Preset]-
>”Printer Driver”->[Printer Supported]).Printer drivers may not be updated in time
due to software version and region limitation, consult printer manufacturer for details.