5-34 Image Optimization
Select [Cur. Win] to be the desired window, then the current active window will be highlighted
by different colors of frame.
A, B, C sectional images (MPR) are corresponding to the following sections of 3D image.
z
Section A: corresponds to the 2D image in B mode. Section A is the sagittal section in
fetal face up posture, as shown in the figure A above.
z
Section B: it is the horizontal section in fetal face up posture, as shown in the figure B
above.
z
Section C: it is the coronal section in fetal face up posture, as shown in the figure C
above.
Tips: the upper part of the 3D image in the D window is corresponding to the orientation mark
on the probe, if the fetal posture is head down (orientating the mother’s feet), and the
orientation mark is orientating the mother’s head, then the fetus posture is head down in the
3D image, you can make the fetus head up by rotating the 3D image by clicking [180°] on the
menu.
Scanning plane and the probe movement
Move the probe across the body surface.
Arrow in the figure below indicates the movement of the probe (you can move the probe
in the contrary direction of the arrow.).
Image Rendering Parameters
Press <Cursor> to show the cursor and roll the trackball to move the cursor onto the target
parameter on the menu, press <Set>/<Back> keys to adjust.
In image viewing status, you can render the image by adjusting the relevant parameters.
Render setting parameters description: