System Preparation 3-1
3
System Preparation
3.1
Move/Posit the System
Please read and understand the safety precautions before placing the system to ensure
safety for both operator and devices.
1. Switch off the power, and pull out the plug.
2. Disconnect the system from all peripherals.
3. Place the system in a desired location by holding the handle.
4. When you move the system to a desired location, lock the four casters.
CAUTION:
1. Maintain enough space around the back and both sides of
the system for plugging/ unplugging the power cord, as
well as avoiding system failure resulted due to increased
system operating temperature.
2. Pay extra attention when moving the system on a sloping
ground, do not move it on a more than 10°-sloped plane to
avoid system toppling.
3.2 Power
Supply
This system can work normally only when it is connected to the external power supply or the
battery capacity is sufficient.
3.2.1
Connecting External Power Supply
A three-wire power cord is used to connect the system with the external power supply.
The external power supply system must meet the following requirements:
Voltage: 220-240V
~
Frequency: 50/60Hz
Input power: 420VA
The connection method is described as follows:
1. Push the retaining clamp upward, and insert the power plug into the receptacle, as shown
in the figure below.