![Shenzhen Mindray Bio-Medical Electronics Co., Ltd. DC-N2 Operator'S Manual Download Page 15](http://html1.mh-extra.com/html/shenzhen-mindray-bio-medical-electronics-co-ltd/dc-n2/dc-n2_operators-manual_1256057015.webp)
Safety Precautions 1-1
1
Safety Precautions
1.1 Safety
Classification
According to the type of protection against electric shock:
CLASS I EQU Internally Powered ME Equipment
According to the degree of protection against electric shock:
Type-BF applied part
According to the degree of protection against harmful ingress of water:
Main unit: IPX0
Probes: IPX7
Footswitch: 1-pedal type belongs to IPX8, 2-pedal and 3-pedal type belong to IP68.
According to the degree of safety of application in the presence of a FLAMMABLE
ANESTHETIC MIXTURE WITH AIR or WITH OXYGEN OR NITROUS OXIDE:
EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANESTHETIC
MIXTURE WITH AIR or WITH OXYGEN OR NITROUS OXIDE
According to the mode of operation:
CONTINUOUS OPERATION
According to the installation and use:
MOBILE EQUIPMENT