12-10 Probes and Biopsy
Manufacturer
Trade Name
Procedures
Type
Minntech
Corporation
Minncare® Cold
Sterilant
Please refer to the instructions
provided by the manufacturer of
the solution for details.
Solution***
“
**” denotes if the disinfectant is not applicable for the probe 65EC10ED.
“
***” denotes if the disinfectant is not applicable for the probe 65EC10ED.
12.1.6 Storage and Transportation
When all examinations for the day have been completed, confirm that the transducer is in
good condition. After disinfecting the transducer, confirm that the transducer is in good
condition and store it in a suitable place so that the next examination can be conducted
smoothly.
1. To prevent the transducer from being damaged, do not store it in locations where it may
be exposed to:
z
Direct sunlight or X-rays
z
Sudden changes in temperature
z
Dust
z
Excessive vibration
z
Heat generators
2. Store and transport the transducer under following ambient conditions:
z
Ambient temperature: -20°C ~ 55°C
z
Relative humidity: 30% to 95% (no condensation)
z
Atmospheric pressure: 700 hPa ~ 1060 hPa
3. When the transducer is sent to MINDRAY Customer Service Department or sales
representative for repair. Please disinfect it and send it in the carrying case.