![Shenzhen Mindray Bio-Medical Electronics Co., Ltd. DC-N2 Operator'S Manual Download Page 210](http://html1.mh-extra.com/html/shenzhen-mindray-bio-medical-electronics-co-ltd/dc-n2/dc-n2_operators-manual_1256057210.webp)
16-4 System Maintenance
Cleaning the peripherals
Do the cleaning maintenance according to your actual peripheral configuration; items which
are not configured can be skipped.
Content
Description
Color and B/W
video printer
First wipe off dust or stain attached to the cover of printer with soft dry
cloth, then clean the inside of printer. Be sure to do the cleaning
maintenance according to the operation manual if is necessary.
Graph / text
printer
First wipe off dust or stain attached to the cover of printer with soft dry
cloth, then clean the inside of printer. Be sure to do the cleaning
maintenance according to the operation manual if is necessary.
Footswitch
Use soft dry cloth with a little mild soap water to wipe off the dust or stain
attached to the pedals or cable of foot switch.
16.1.2 Checking the Transducer
Visually check to confirm that there is no crack and expansion to transducer head.
Visually check to confirm that there is no deterioration or desquamation to probe cable.
Visually check to confirm that there is no bend, destroyed or falling off pins to the
connector.
16.1.3 Checking the Power Cable and Plug
Visually check to confirm that there is no wrinkles, crack or deterioration; No crack and
expansion to the surface of adapter.
Manually check to confirm that there is no looseness or rupture. The connection of plug is
reliable.
16.1.4 Checking the Circuit Breaker
Manually check to confirm that the circuit breaker can be turned on or off normally.
16.1.5 Checking
Appearance
Check if there is any crack in the covers and:
Caster
1. Visually check to confirm there is no any crack.
2. Operate the casters to confirm the locking and releasing functions
are normal.
Connection of the
caster
1. Visually check to confirm that there is no skewness and the
connecting screws are free of breakage or falling off.
2. Check with the spanner to make sure that there is no looseness
between the caster and the base connection screw.
Handle
1. Visually check to confirm there is no any crack.
2. Check by hand to confirm that the handle is free of looseness.
Lifting structure
Hold the lifting arm control handle, make sure that the lifting arm can go
up and down normally without abnormal sounds or phenomena
Control panel
support assembly
Check by hand to confirm that the control panel is free of looseness.