4. STAT
IM 5000 Calibration
2. validation thermocouple calibration
To calibrate the validation thermocouple:
1. Turn the unit off
oFF
.
2. Turn ON the unit while keeping Unwrapped and Wrapped button pressed to enter STAT
IM
Service Mode
.
3. This
Service Mode
is password protected, enter password to continue (default password
is: Unwrapped, Wrapped, R&P and Stop keys pressed in this order).
4. Toggle using keypad through the menu selection to reach Calibration option and press
R&P key.
5. The display should show the calibration screen as before
6. To start a validation thermocouple (self-calibrating) cycle,
press and hold
the
UNWRAPPED cycle button on the keypad/control box and press the
START
button. An
asterisk “*” will appear next to the validation thermocouple offset value (CVO) to indicate
that a Validation Assy. calibration cycle is running.
7. If the asterisk “*” does not appear, press the
SToP
button on the control box twice to
reset the unit. Turn the power switch
oFF
then
oN
again.
8. Hold down the UNWRAPPED cycle button on the control box and press the
START
button
to start the self-calibration again.
9. This calibration will take approximately 6 minutes.
10. Allow the self-calibration to complete.
11. The temperature within the chamber will rise to the sterilization temperature. Wait until
sterilization phase of the calibration cycle ends automatically.
12. The offset value in the upper left-hand corner of the display (CVO) may have changed to a
new offset value.
13. Press the
SToP
button on the control box when finished.
3.
Switch off the unit and remove all calibration equipment.