![SciCan STATIM 5000S Service Manual Download Page 7](http://html1.mh-extra.com/html/scican/statim-5000s/statim-5000s_service-manual_1217148007.webp)
Page 7
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
1. Identifying STAT
IM 5000 Units
STAT
IM
5000/5000S/5000 G4 Service Guide
Identifying STAT
IM 5000 Units
There are two main types of STAT
IM 5000 units, the G4 touchscreen version and STATIM classic
with keypad. Each of these two versions have a North American specification or “non S” unit,
and the European specification or “S” unit. In the classic series, they are labeled respectively as
“STAT
IM 5000” and “STATIM 5000S”. Though it does not use the “S” designation in its model
names, the G4 series has S-class units for distribution in markets requiring these cycle types, and
non S-class for the North American market.
There are many mechanical identifiers to differentiate between the set types, (e.g., pressure
transducer for ‘S’ class, no pressure transducer for the non ‘S’ class). The details in the matrix
further below will help you identify units based on their factory specifications.
Identifying Software versions
To identify a unit’s software version, power it OFF and back ON again. During start up, a
number is briefly displayed in the upper right-hand corner of the LCD that indicates which
software is installed (See Figure 1). This number is also printed on the label attached to the U14
microprocessor and the U8EEPROM device on the controller board (See Figure 2).
The software version number is the character to the immediate right of the R as indicated by:
Figure 1
S501R102
S501R200
example
example
LCD readout on Power-up
S501R100 is software version 1.00
S501R101 is software version 1.01
S501R102 is software version 1.02
S501R200 is software version 2.00
For example:
Figure 2
01-xxxxxx
S502R102
Sept. XX 03
52102
0319
Microprocessor label example:
EPROM label
example:
Product Year / week