Setting Air Drying Time
NOTE:
This function is available on Rev. 7 controller boards with software revision SXXXR612 on
STAT
im
Classic, and software revision SXXXR700 on STAT
im
G4. However, this function is not
available on U.S. STAT
im
models.
For STAT
im
Classic, follow these steps:
1. Power up unit while pressing the
STOP
button to access the User Setup menu.
2. Using the UNWRAPPED and WRAPPED buttons, scroll to AIR DRYING and select it by
pressing the RUBBER AND PLASTIC button.
3. Use the UNWRAPPED and WRAPPED buttons to change the displayed values and the
RUBBER AND PLASTIC button to select and move to the next digit. Press the
STOP
button
when completed.
For STAT
im
G4, follow these steps:
1. Access the Service Menu as described in Using the LCD Service Menu on the STAT
im
G4 in
Chapter 3 and scroll to DRYING TIME.
2. Use the first pair of the arrow buttons to move from one digit to the next. Use the second pair
of arrow buttons to change digit values.
3. To save changes and return to the regular operating mode, press the
STOP
button.