Page 45
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
2. Tools, Maintenance Schedules, Procedures and Testing
STAT
IM
5000/5000S/5000 G4 Service Guide
Water Conductivity Measurement
Figure 27
Water Conductivity Testing
Water conductivity testing determines the amount of
dissolved solids in the steam-process distilled water used
in the STAT
IM unit. Use only steam-process distilled water
having 5 ppm or less dissolved solids or a conductivity of
less than 10 μS / cm. Follow the manufacturer’s instructions
to test water conductivity using SciCan conductivity meter,
part #01-103139S, or any other equivalent Meter.
Temperature has a significant effect on conductivity
readings, therefore the water being tested should be at
room temperature.