Page 122
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
6. Cover Assembly
Removing and Replacing the Armature Gasket
To remove the armature gasket, remove the cover to access the inside of the fascia and peel the
existing armature gasket from the fascia.
To replace the armature gasket proceed as follows (see Figure 2).
1. Remove all traces of the gasket and gasket adhesive from inside the fascia (1). The fascia is
manufactured of polycarbonate plastic. DO NOT USE AROMATIC SOLVENTS ON THE FASCIA.
This will damage the fascia. To prepare for installation of the new gasket (2), roughen the inside
surface of the fascia using a fine grit sand paper.
2. Leave the carrier paper on the adhesive side of the gasket and perform a trial fit. Note that the
widest part of the gasket fits at the bottom of the opening and that the gasket does not protrude
into the opening.
3. Remove the carrier paper from the adhesive side of the gasket.Carefully install the gasket
making sure it is seated firmly in place.
Removing and
Replacing the Fascia
To remove the fascia (1), proceed
as follows (see Figure 2):
1. The fascia is attached to the
cover (3) using seven plastite
screws (4). Remove and retain
the screws.
2. A bead of silicon was used
to provide a water-tight
seal between the facsia and
cover during manufacturing.
Remove the fascia and any
silicon residue remaining on
the cover.
3. If a printer module (5) is installed it must be removed from the fascia. See, 12. Printer -
Removing and Replacing the Printer.
4. Discard the old fascia. The fascia is made of recyclable polycarbonate. Please recycle wherever
possible.
To replace the fascia (1), proceed as follows (see Figure 2):
1. If a printer module (5) was previously installed it must be reinstalled in the fascia. See, Removing
and Replacing the Printer.
2. Apply a bead of silicon between the facsia and cover (3) to provide a water-tight seal.
3. Install the new fascia using the seven screws (4) retained from the disassembly procedure.
4. Reinstall the cover. See, Cover Removal and Replacement, above. Verify that the keypad, LCD
and printer (if installed) function correctly.
1. fascia
2. armature gasket
3. cover
4. plastite screw
5. printer module
Figure 2
STAT
IM
5000/5000S/5000 G4 Service Guide