7. Electrical and Electronic Components
STAT
IM
5000/5000S/5000 G4 Service Guide
Removing a Detachable Power Cord
To remove a detachable power cord (1) connected to the unit through a panel mount A.C. inlet
receptacle, proceed as follows (see Figure 20):
1. Turn the power switch (2) OFF, and unplug the power cord from the wall outlet.
2. Disconnect the power cord from the panel mount A.C. inlet receptacle (3).
Replacing/refitting a Detachable Power Cord
use replacement cords with appropriate safety agency ratings and approvals only.
To replace/reinstall a detachable power cord (1), proceed as follows (see Figure 20):
1. Select the appropriate replacement cord from the spare parts list.
2. The A.C. inlet receptacle (3) is keyed. Note the shape of the cord connector and the
corresponding shape of the receptacle. Plug in the connector.
3. Plug the cord into the wall receptacle and turn the power switch (2) ON. Observe the LCD and
indicator lights to determine that power is present.