![SciCan STATIM 5000S Service Manual Download Page 69](http://html1.mh-extra.com/html/scican/statim-5000s/statim-5000s_service-manual_1217148069.webp)
Page 69
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
3. Troubleshooting Cycle Faults
STAT
IM
5000/5000S/5000 G4 Service Guide
C
F
N
o.
U
ni
t
typ
e
So
ftw
ar
e
re
vi
si
on
D
esc
rip
tion
o
f F
au
lt
Pr
ob
ab
le
C
au
se
(o
r causes
) o
f F
au
lt
S
N
on
S
1.
xx
/
2.
xx
4.
xx
5.
xx
6.
xx
5
1.xx only
The
s
oft
w
ar
e
cau
se
s
the
pu
m
p
to
a
cti
va
te
a
t
pr
ede
te
rm
ined
ti
m
es
. If
a
reque
st t
o
pu
m
p
w
at
er
occ
ur
s
ou
ts
ide
o
f t
he
p
rede
te
rm
ined
ti
m
e
C
yc
le
Fau
lt
#5
o
cc
ur
s.
•
Lea
ky
c
ass
ett
e
•
K
in
ked
pu
m
p t
ub
e
•
Lea
ki
ng
s
ol
eno
id
v
al
ve
•
P
oo
r pu
m
p
de
live
ry
•
O
ve
rloaded
c
ass
ett
e
6 (i)
2.xx only
The
s
oft
w
ar
e
ha
s
de
te
ct
ed
the
fil
te
red
s
tea
m
gene
ra
to
r t
em
pe
ra
tu
re
6
°C
g
rea
te
r t
han
the
cha
m
be
r du
ring
the
s
te
rili
zi
ng
phase
o
f a
c
yc
le
.
•
O
bs
tru
cti
on
c
au
sed
b
y:
o
S
ea
l m
is
ali
gn
m
en
t
o
Fo
re
ign
m
at
er
ia
l i
n t
he
s
ea
l
channe
l
•
C
ali
br
at
ion
im
ba
lan
ce
(bo
ile
r
the
rm
oc
oup
le
to
c
ha
m
be
r
the
rm
oc
oup
le
)
•
C
on
ta
m
ina
ted
s
tea
m
gene
ra
to
r
6 (ii)
The
s
oft
w
ar
e
ha
s
de
te
ct
ed
the
V
ali
da
tion
the
rm
oc
oup
le
te
m
pe
ra
tu
re
to
be
5
°C
g
rea
te
r
than
the
c
ha
m
be
r du
ring
the
s
te
rili
zi
ng
pha
se
o
f
a
cyc
le
.
•
C
ass
ett
e
st
ea
m
lea
k
•
S
tick
ing
s
ol
eno
id
•
O
bs
tru
ct
ed
e
xhau
st t
ub
ing
(k
in
k)
•
C
ali
br
at
io
n
im
ba
lan
ce
(va
lida
tion
the
rm
oc
oup
le
to
c
ha
m
be
r
the
rm
oc
oup
le
)
C
he
ck
V
re
f i
s
2.
520
V
±
0.
001
V
(a
ll
so
ftw
ar
e
re
vi
si
on
s
O
TH
E
R
T
H
A
N
re
vi
si
on
6
.xx
)
7
D
ur
in
g
the
s
te
rili
za
tion
pha
se
o
f t
he
cyc
le
the
casse
tte
te
m
pe
ra
tu
re
has
d
ropped
be
lo
w
a
th
re
sho
ld
v
al
ue
.
•
S
tick
ing
s
ol
eno
id
va
lve
(open
)
•
C
ass
ett
e
lea
k
•
C
he
ck
va
lve
lea
k
•
P
R
V
lea
k