Page 251
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
12. Printer and Data Logger
STAT
IM
5000/5000S/5000 G4 Service Guide
11. Using the UNWRAPPED and WRAPPED buttons, scroll until reaching the desired Hex value
in the table above (or from the printer’s user manual) and press the RUBBER & PLASTICS
button to select and return to the user menu.
Important note: the UNWRAPPED button will
increase the displayed value by 1, and the WRAPPED button will increase that displayed
value by ten.
12. Using the UNWRAPPED and WRAPPED buttons, scroll until the cursor is next to “Save and
Exit” and press the RUBBER & PLASTICS button, which will now save the data, and take you
out of the user menu and back to the operating menu.
Note: When printed or electronic records are being kept for future reference, it is important
to ensure the correct data is printed, and a check should be undertaken at this stage to
ensure that the date, time and unit number (where appropriate) are set before using a data
recording device. This may have been undertaken during installation, if not, refer to the user
manual or to this service manual’s Chapter 13, “Using the service menu” for information on
resetting the time, date and unit number.
13. Power OFF the STAT
im
unit.
Connecting the Printer
1. Connect the printer to the STAT
im
unit using a Serial Null Modem Cable DB9F – DB25M and
secure the cable.
2. Connect the printer power supply to the mains power socket as appropriate to the printer
being used.
3. Power ON the printer.
4. Power ON the STAT
im
unit.
5. The unit is now ready for use.
Data logger
(for use on Rev. 7.x controller boards oNlY)
Installing the SciCan Data logger
Setting the STAT
IM
unit
SciCan’s Data Logger can record and store cycle information onto a mass storage device (MSD)
such as a USB Flash Drive or SD memory card.
The Data Logger is supplied with two cables for connection to 25 pin (early STAT
im
2000 units
only) and 9 pin (all revision 7 2000 and 5000 units) data ports. The 9 pin cable is the only one
required for 5000 units.
NOTE: Statim 5000 units manufactured with internal printers will NOT have a communication
port fitted. However,
• Units that were manufactured with Revision 7.x controller boards should have a
cut-out in the chassis with a blanking plate and can be upgraded for data logger
use by installing RS232 Port Kit Statim 5000/5000S part number 01-110222S.