Measurement display
Activating the measurement function, the instrument displays the total
intermodulation product.
Frequency spectrum
1.
Call the measurement function
POST
FFT
(
Ê
2.
Select the
SPECTRUM
display mode in the Graph menu (
Ê
A measurement diagram with the frequency spectrum is shown in the
display area.
Note:
In the Graph menu you can change the graphic display area
(
Ê
6-292, 6-296) and analyze the trace using the cursors (
Ê
Bar graph
Select the
BAR GRAPH
display mode in the Graph menu (
Ê
A bar graph is shown in the display area. Depending on the measure-
ment type, the interference signal (1), useful signal (4), and
intermodulation products (2, 3, 5, 6) of the input signal are
displayed.
Note:
When you activate the measurement function, the R&S UP300/350
begins the continuous measurement. However, you can also start and stop
the measurement manually (
Ê
Summary of Contents for Smart Instruments UP300