![Rohde & Schwarz Smart Instruments UP300 Operating Manual Download Page 157](http://html1.mh-extra.com/html/rohde-and-schwarz/smart-instruments-up300/smart-instruments-up300_operating-manual_1477754157.webp)
R&S UP300/350
Generator
Operating Manual, 11/2007
6-157
E-1147.2759.00
Selecting the
sweep mode
1.
Press the
function key
in the
menu.
A selection field containing the available settings is displayed. The
default setting is “FREQ SWEEP “.
2.
Use the
rotary knob
[11] to select a setting.
3.
Press the
ENTER key
[5] to close the selection field.
The new setting is stored and displayed in the display area.
Setting the Measurement Time
Use
You can set the measurement time dependent on the task.
Note:
The setting is always valid for both channels (Ch 1&2).
Setting the
measurement time
1.
Press the
function key
in the
menu.
An entry field containing the currently applicable setting is displayed. The
default setting is “10 ms“. At the same time, the function keys [13] are
assigned various units of measurement.
2.
Enter a new value (
Ê
The permissible entry range is:
1 ms
≤
MEAS TIME
≤
10 s
The new setting is stored and displayed in the parameter field.
Ch 1&2
Ch 1&2
Summary of Contents for Smart Instruments UP300
Page 1: ...11 2007...
Page 20: ...R S UP300 350 E 1147 2759 00 1 20 Operating Manual 11 2007...
Page 30: ......