R&S UP300/350
Analyzer
Operating Manual, 11/2007
6-273
E-1147.2759.00
Selecting the Type of Signal Search
Note:
The lower limit frequency (f
min
)
of the measurement signal depends on
the selected bandwidth of the analyzer.
Bandwidth 22 kHz:
f
min
≥
20 Hz
Bandwidth 40 kHz:
f
min
≥
40 Hz
Bandwidth 80 kHz:
f
min
≥
80 Hz
Use
Depending on the measurement task, you can select different modes for the
signal frequency search in channel Ch 1:
Auto tuning
Automatic signal frequency search and automatic phase measurement
Fixed frequency
Manual entry of signal frequency and manual phase measurement
Note:
The setting is always valid for both channels (Ch 1&2).
Selecting the type of
signal search
1.
Press the
function key
in the
menu.
A selection field containing the available settings is displayed. The
default setting is “Auto tuning“.
2.
Use the
rotary knob
[11] to select a setting.
3.
Press the
ENTER key
[5] to close the selection field.
The new setting is stored and displayed in the parameter field.