Generator R&S
UP300/350
E-1147.2759.00
6-198
Operating Manual, 11/2007
Entering the
stop value
3.
Press the
function key
in the
submenu
.
An entry field containing the currently applicable setting is displayed. The
default setting is “22.139 kHz“. At the same time, the function keys [13]
are assigned various units of measurement.
4.
Enter a new value (
Ê
The permissible entry range is:
10 Hz
≤
Stop
≤
f
max
within: f
max
- maximum frequency of generator type (
Ê
The new setting is stored and displayed in the parameter field.
Note:
If the start value is higher than the stop value, the start value is
adjusted automatically.
Selecting the Spacing of Sweep Steps (Lin/Log)
Use
The spacing of the sweep points can be selected. You can decide if the
distance between the valus set one after another is in
Linear
or
Logarithmic
steps.
Selecting the
Spacing
1.
Press the
function key
in the corresponding
or
submenu
.
A selection field containing the available settings is displayed. The
default setting is “Linear“.
2.
Use the
rotary knob
[11] to select a setting.
3.
Press the
ENTER
key [5] to close the selection field.
The new setting is stored and displayed in the parameter field.