![Rohde & Schwarz Smart Instruments UP300 Operating Manual Download Page 102](http://html1.mh-extra.com/html/rohde-and-schwarz/smart-instruments-up300/smart-instruments-up300_operating-manual_1477754102.webp)
Generator R&S
UP300/350
E-1147.2759.00
6-102
Operating Manual, 11/2007
6.2.1.2 Analog
Generator
6.2.1.2.1
Selecting the Generator Bandwidth
Use
Switching the bandwidth changes the sample rate of the signal. Since the
properties of the digital filters have become less favourable as the sample
rate increases, you should select the lowest possible bandwidth for your
specific application.
The R&S UP300/350 provides the following bandwidths for the analog
generator:
22 kHz
40 kHz
80 kHz
Note:
The setting is always valid for both channels (Ch 1&2).
Selecting the
bandwidth
3.
Press the
function key
in the
menu.
A selection field containing the available settings is displayed. The
default setting is “22 kHz “.
4.
Use the
rotary knob
[11] to select a setting.
5.
Press the
ENTER key
[5] to close the selection field.
The new setting is stored and displayed in the parameter field.
Ch 1&2