R&S UP300/350
Analyzer
Operating Manual, 11/2007
6-227
E-1147.2759.00
6.3.2.2
Configuring the Measurement Parameters
Use
In the FUNCTIONS menu (
Ê
6-226), you can select the analyzer function.
The selected measurement function is displayed in a vacant field in the menu
bar and is available as a menu key to allow you to enter parameters for the
function. At the same time, the appropriate function is activated in the
analyzer.
When you have selected a measurement function in the menu bar, the list of
available function parameters (e.g. FREQ., DC, RMS) appears on the screen.
Settings for the
selected channel
Some parameters of the analyzer functions (FILTER, CONFIG) can be set
channel independently (Ch 1, Ch 2), or simultaneously (Ch 1& 2). These
parameters are listed in two columns in the parameter field. The function
parameters applying to both channels (Ch 1&2) are listed in one column.
To get an overview, refer to the factory default settings (
Ê
Reference unit
All measurements which return results with dimensions can be displayed
either as absolute measurements or relative to a reference value. If you
select the reference unit (dBr, dBrFS), the measurement result is displayed
taking the entered reference value into consideration. Each measurement
function has its own reference value that can be set separately for both
channels.