Command Reference
R&S
®
CMW-KM750/-KM751
67
User Manual 1173.9670.02 ─ 05
..........................................................70
....................................71
.........................................................71
...................................................72
....................................72
CONFigure:TDSCdma:MEAS<i>:UESignal:FAUSlot
<FstActUpSlot>
Specifies the first active uplink timeslot in the subframe (relevant for IF trigger source).
Parameters:
<FstActUpSlot>
SL1 | SL2 | SL3 | SL4 | SL5 | SL6
Timeslot 1 ... 6
*RST:
SL1
Example:
See
Firmware/Software:
V1.0.4.11
Manual operation:
See
CONFigure:TDSCdma:MEAS<i>:UESignal:CDTHreshold
<ChnDetThres>
Selects the threshold for automatic channelization code search. The threshold deter-
mines whether a signal is evaluated or rejected (noise, non-orthogonal components).
Parameters:
<ChnDetThres>
Range:
-128 dB to 127 dB
*RST:
-20 dB
Default unit: dB
Example:
See
Firmware/Software:
V1.0.4.11
Manual operation:
See
CONFigure:TDSCdma:MEAS<i>:UESignal:SPOint
<SwitchPoint>
Selects the position of the duplex switching point in the subframe.
Parameters:
<SwitchPoint>
TS1 | TS2 | TS3 | TS4 | TS5 | TS6
Duplex switching point is located after the selected timeslot, e.g.
TS2 means after timeslot 2
*RST:
TS3
Example:
See
Firmware/Software:
V1.0.4.11
Manual operation:
See
General Measurement Settings