4.2.1 Signal Routing and Analyzer Settings
The following parameters configure the RF input path.
See also: "Connection Control (Measurements)" in the R&S CMW user manual, chapter
"System Overview"
Fig. 4-1: Signal routing and analyzer settings
Scenario
This software version supports only a standalone scenario.
Remote command:
RF Routing
Selects the input path for the measured RF signal, i.e. the input connector and the RX
module to be used.
Depending on your hardware configuration there may be dependencies between both
parameters. Select the RF connector first. The "Converter" parameter offers only values
compatible with the selected RF connector.
Remote command:
External Attenuation (Input)
Defines the value of an external attenuation (or gain, if the value is negative) in the input
path. The power readings of the R&S
CMW are corrected by the external attenuation
value.
The external attenuation value is also used in the calculation of the maximum input power
that the R&S
CMW can measure.