Command Reference
R&S
®
CMW-KM750/-KM751
103
User Manual 1173.9670.02 ─ 05
<EMRelBLimit>
Limit at point B
Range:
-100 dB to 0 dB
*RST:
-49 dB
Default unit: dB
Additional parameters: OFF | ON (disables | enables the limit
check using the previous/default limit values)
<EMRelCLimit>
Limit at point C
Range:
-100 dB to 0 dB
*RST:
-64 dB
Default unit: dB
Additional parameters: OFF | ON (disables | enables the limit
check using the previous/default limit values)
<EMRelDLimit>
Limit at point D
Range:
-100 dB to 0 dB
*RST:
-49 dB
Default unit: dB
Additional parameters: OFF | ON (disables | enables the limit
check using the previous/default limit values)
<EMRelELimit>
Limit at point E
Range:
-100 dB to 0 dB
*RST:
-35 dB
Default unit: dB
Additional parameters: OFF | ON (disables | enables the limit
check using the previous/default limit values)
Example:
See
Firmware/Software:
V1.0.4.11
6.3.10 EVM Results (Traces)
The following commands return the EVM trace results of the multi evaluation measure-
ment.
Multi Evaluation Measurement Commands