Programming
R&S
®
CMW-KM750/-KM751
53
User Manual 1173.9670.02 ─ 05
5.1.5 Specifying Limits
// *****************************************************************************
// Define modulation limits
// *****************************************************************************
CONFigure:TDSCdma:MEAS:MEValuation:LIMit:MERRor 20, OFF
CONFigure:TDSCdma:MEAS:MEValuation:LIMit:EVMagnitude 10, 20
CONFigure:TDSCdma:MEAS:MEValuation:LIMit:PERRor 20, OFF
CONFigure:TDSCdma:MEAS:MEValuation:LIMit:IQOFfset -20
CONFigure:TDSCdma:MEAS:MEValuation:LIMit:IQIMbalance ON
CONFigure:TDSCdma:MEAS:MEValuation:LIMit:FERRor 150
CONFigure:TDSCdma:MEAS:MEValuation:LIMit:TERRor 2
CONFigure:TDSCdma:MEAS:MEValuation:LIMit:PCDerror -10
CONFigure:TDSCdma:MEAS:MEValuation:LIMit:WQUality 0.95
// *****************************************************************************
// Define power vs. time limits
// *****************************************************************************
CONFigure:TDSCdma:MEAS:MEValuation:LIMit:PVTime:YABSolute -70,-60,-70
CONFigure:TDSCdma:MEAS:MEValuation:LIMit:PVTime:XRELative -55,-33,-13,12,45
CONFigure:TDSCdma:MEAS:MEValuation:LIMit:PVTime:EAReas ON,ON,ON
// *****************************************************************************
// Define ACLR limits
// *****************************************************************************
CONFigure:TDSCdma:MEAS:MEValuation:LIMit:ACLR:OBW 1.4
CONFigure:TDSCdma:MEAS:MEValuation:LIMit:ACLR:YABSolute ON
CONFigure:TDSCdma:MEAS:MEValuation:LIMit:ACLR:YRELative -33, -43
// *****************************************************************************
// Define spectrum emission mask
// *****************************************************************************
CONFigure:TDSCdma:MEAS:MEValuation:LIMit:EMASk:YABSolute -55.8,-71.1
CONFigure:TDSCdma:MEAS:MEValuation:LIMit:EMASk:YRELative -42,-42,-57,-47,-33.1
5.1.6 Performing Single-Shot Measurements
// *****************************************************************************
// Enable all measurements and select a code channel for the CDP
// multislot results.
// *****************************************************************************
CONFigure:TDSCdma:MEAS:MEValuation:RESult ON
CONFigure:TDSCdma:MEAS:MEValuation:CDPower:MSLot:CINDex 10
// *****************************************************************************
// Start single-shot measurement, return magnitude error trace.
// Query the measurement state (should be "RDY").
// *****************************************************************************
READ:TDSCdma:MEAS:MEValuation:TRACe:MERRor:AVERage?
General Examples