Command Reference
R&S
®
CMW-KM750/-KM751
85
User Manual 1173.9670.02 ─ 05
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V1.0.2.7
Manual operation:
See
CONFigure:TDSCdma:MEAS<i>:MEValuation:SCOunt:MODulation
<StatCntMod>
Specifies the statistic count of the measurement. The statistic count is equal to the num-
ber of measurement intervals per single shot. Use
CONFigure:...:MEAS<i>:...:REPetition SINGleshot | CONTinuous
to
select either single-shot or continuous measurements.
See also: "Statistical Settings" in the R&S CMW user manual, chapter "Remote Control"
Parameters:
<StatCntMod>
Number of measurement intervals.
Range:
1 to 1000
*RST:
10
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V1.0.2.7
Manual operation:
See
"Power vs Time / Modulation / Spectrum / BER > Statistic
CONFigure:TDSCdma:MEAS<i>:MEValuation:SCOunt:PVTime
<StatCntPvT>
Specifies the statistic count of the measurement. The statistic count is equal to the num-
ber of measurement intervals per single shot. Use
CONFigure:...:MEAS<i>:...:REPetition SINGleshot | CONTinuous
to
select either single-shot or continuous measurements.
See also: "Statistical Settings" in the R&S CMW user manual, chapter "Remote Control"
Parameters:
<StatCntPvT>
Number of measurement intervals.
Range:
1 to 1000
*RST:
10
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V1.0.2.7
Manual operation:
See
"Power vs Time / Modulation / Spectrum / BER > Statistic
CONFigure:TDSCdma:MEAS<i>:MEValuation:HDALevel
<HighDynAsemLev>
Defines a signal level relative to the "Expected Nominal Power" (see
on page 65) where the two results
obtained in a two stage measurement are joined together.
Multi Evaluation Measurement Commands