ABORt...
causes a running measurement to stop immediately; the R&S
CMW
enters the "OFF" state.
Use
FETCh...STATe?
to query the current measurement state.
See also: "Measurement Control" in the R&S CMW user manual, chapter "Remote Con-
trol"
Example:
See
Single-Shot and Continuous Measurements
Usage:
Event
Firmware/Software:
V1.0.2.7
Manual operation:
See
FETCh:TDSCdma:MEAS<i>:MEValuation:STATe?
Queries the main measurement state. Use
FETCh:...:STATe:ALL?
to query the mea-
surement state including the substates. Use
INITiate..., STOP..., ABORt...
to change the measurement state.
See also: "Measurement Control" in the R&S CMW user manual, chapter "Remote Con-
trol"
Return values:
<MeasState>
OFF | RUN | RDY
OFF:
measurement switched off, no resources allocated, no
results available (when entered after
ABORt...
)
RUN:
measurement running (after
INITiate..., READ...
),
synchronization pending or adjusted, resources active or queued
RDY:
measurement has been terminated, valid results may be
available
*RST:
OFF
Example:
See
Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V1.0.2.7
Manual operation:
See
FETCh:TDSCdma:MEAS<i>:MEValuation:STATe:ALL?
Queries the main measurement state and the measurement substates. Both measure-
ment substates are relevant for running measurements only. Use
FETCh:...:STATe?
to query the main measurement state only. Use
INITiate...,
STOP..., ABORt...
to change the measurement state.
See also: "Measurement Control" in the R&S CMW user manual, chapter "Remote Con-
trol"