Command Reference
R&S
®
CMW-KM750/-KM751
143
User Manual 1173.9670.02 ─ 05
Return values:
<1_Reliability>
In list mode, a zero reliability indicator indicates that the results in
all measured segments are valid. A non-zero value indicates that
an error occurred in at least one of the measured segments.
{<2_SegReliablity>
Reliability indicator for the segment. The meaning of the returned
values is the same as for the common reliability indicator, see
previous parameter.
<3_UEpower>
User equipment power
Range:
-100 dBm to 35 dBm
Default unit: dBm
<4_CarrierPower>
Power at the nominal carrier frequency (UL Frequency)
Range:
-100 dBm to 35 dBm
Default unit: dBm
<5_ACLRMinus2>
Power of adjacent channel at -3.2 MHz from UL frequency
Range:
-100 dBm to 35 dBm
Default unit: dBm
<6_ACLRMinus1>
Power of adjacent channel at -1.6 MHz from UL frequency
Range:
-100 dBm to 35 dBm
Default unit: dBm
<7_ACLRPlus2>
Power of adjacent channel at +3.2 MHz from UL frequency
Range:
-100 dBm to 35 dBm
Default unit: dBm
<8_ACLRPlus1>
Power of adjacent channel at +1.6 MHz from UL frequency
Range:
-100 dBm to 35 dBm
Default unit: dBm
<9_OBW>
Occupied bandwidth
Range:
0 MHz to 4 MHz
Default unit: Hz
<10_MarginAB>
<11_MarginCD>
<12_MarginDE>
<13_MarginED>
<14_MarginDC>
<15_MarginBA>}
Limit line margin values in the 6 emission mask areas
Range:
-100 dB to 100 dB
Default unit: dB
Example:
See
Usage:
Query only
Firmware/Software:
V3.0.10
Options:
R&S CMW-KM012
Multi Evaluation Measurement Commands