Command Reference
R&S
®
CMW-KM750/-KM751
90
User Manual 1173.9670.02 ─ 05
<Band>
B1 | B2 | B3
B1:
Band 1 (F), 1880 MHz to 1920 MHz
B2:
Band 2 (A), 2010 MHz to 2025 MHz
B3:
Band 3 (E), 2300 MHz to 2400 MHz
*RST:
B1
<Retrigger>
OFF | ON
Specifies whether a trigger event is required for the segment or
not. The setting is ignored for the first segment of a measurement
and for trigger mode ONCE (see
OFF
: measure the segment without retrigger
ON
: wait for trigger event before measuring the segment
*RST:
OFF
Example:
See
Firmware/Software:
V3.0.10
V3.0.20: added <Level>, <Frequency>, <Band> and <Retrigger>
V3.0.30: max. no. of segments increased to 500 / subframes
increased to 1600
Options:
R&S CMW-KM751 required for modulation type EPSK and QAM
R&S CMW-KM012
CONFigure:TDSCdma:MEAS<i>:MEValuation:LIST:SEGMent<Segno>:
MODulation
<ModStatistics>, <EnableEVM>, <EnableMagError>,
<EnablePhaseError>, <EnableIQ>, <EnableModScalar>, <EnablePMonitor>
Defines the statistical length for the
AVERage
,
MAXimum
, and
SDEViation
calculation
and enables the calculation of the different modulation results in segment <Segno>.
Suffix:
<Segno>
.
1..500
Parameters:
<ModStatistics>
Statistical length in slots. The maximum possible value depends
on the number of subframes in the segment and on the number of
active uplink slots per subframe, see
on page 88.
Range:
1 to 1000
*RST:
10
<EnableEVM>
OFF | ON
OFF:
Disable measurement of EVM
ON:
Enable measurement of EVM
*RST:
ON
Multi Evaluation Measurement Commands