3.5 Limit Settings and Conformance Requirements
Conformance requirements for TD-SCDMA transmitter tests are specified in 3GPP TS
34.122, section 5, "Transmitter Characteristics".
The following sections give an overview of the R&S
CMW limit settings and the related
test requirements.
.....................................................................................20
........................................................................................23
3.5.1 Transmit Modulation Limits
A poor modulation accuracy of the UE transmitter increases the transmission errors in
the uplink channel of the TD-SCDMA network. The Error Vector Magnitude (EVM) is the
critical quantity to assess the modulation accuracy of a TD-SCDMA UE.
According to 3GPP, the EVM measured at UE output powers
≥
−
20 dBm and under
normal operating conditions shall not exceed 17.5 %. In addition, the Peak Code Domain
Error, which is the maximum EVM of all code channels, shall not exceed –20 dB. The
frequency error shall not exceed ±0.1 ppm. These limits can be set in the configuration
dialog, along with limits for the other measured quantities.