Command Reference
R&S
®
CMW-KM750/-KM751
84
User Manual 1173.9670.02 ─ 05
Parameters:
<ScalarSlot>
Range:
0 to 7 * <measured subframes> - 1
*RST:
1
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V1.0.4.11
Manual operation:
See
CONFigure:TDSCdma:MEAS<i>:MEValuation:SFOFfset
<SubframeOffset>
Specifies a subframe number mismatch to be compensated for RX measurements. Set
the number of the subframe that contains the looped back data expected in subframe
number 0.
Parameters:
<SubframeOffset>
Range:
0 to 3
*RST:
3
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V3.0.10
Manual operation:
See
CONFigure:TDSCdma:MEAS<i>:MEValuation:AMODe
<AnaMode>
Defines whether a possible origin offset is included in the measurement results (WOOFf-
set) or subtracted out (NOOFfset).
Parameters:
<AnaMode>
NOOFfset | WOOFfset
NOOFfset
: No origin offset
WOOFfset
: With origin offset
*RST:
NOOF
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V1.0.2.7
Manual operation:
See
CONFigure:TDSCdma:MEAS<i>:MEValuation:MOEXception
<MeasOnExp>
Specifies whether measurement results that the R&S
CMW identifies as faulty or inac-
curate are rejected.
Parameters:
<MeasOnExp>
OFF | ON
OFF
: Faulty results are rejected.
ON
: Results are never rejected.
*RST:
OFF
Multi Evaluation Measurement Commands