82
Commands
Data output related commands
DOD
Output displayed value (no corresponding operation)
If the unit is displaying a menu or is in recall mode, the command is
not valid and error 0003 is returned.
Send the request command at one second interval or longer.
Request command
DOD ?
<Sound level meter mode>
Response data
d1,d2,...,d23
d1 = xxx.x
Main channel
L
p
d2 = xxx.x
Main channel
L
eq
d3 = xxx.x
Main channel
L
E
d4 = xxx.x
Main channel
L
max
d5 = xxx.x
Main channel
L
min
d6 = xxx.x
Main channel
L
N
1
d7 = xxx.x
Main channel
L
N
2
d8 = xxx.x
Main channel
L
N
3
d9 = xxx.x
Main channel
L
N
4
d10 = xxx.x
Main channel
L
N
5
d11 = xxx.x
Sub channel
L
p
d12 = xxx.x
Sub channel
L
eq
d13 = xxx.x
Sub channel
L
E
d14 = xxx.x
Sub channel
L
max
d15 = xxx.x
Sub channel
L
min
d16 = xxx.x
Sub channel
L
N
1
d17 = xxx.x
Sub channel
L
N
2
d18 = xxx.x
Sub channel
L
N
3
d19 = xxx.x
Sub channel
L
N
4
d20 = xxx.x
Sub channel
L
N
5
d21 = xxx.x
Sub channel
L
peak
/
L
tm5
d22 = 0 or 1
Overload information (1: Yes, 0: No)
d23 = 0 or 1
Under-range information (1: Yes, 0: No)
* d1 to d21 are fi xed at 5 digit length. Higher digits are padded
with spaces as required.
* When display is set to OFF, d1 to d21 are returned as “_--.-”
(with a leading space).