58
Commands
LXI
Percentile level (corresponds to Display
L
N
1
to
L
N
5
setting
value)
Only valid if the unit is currently in
L
p
mode (except pause). In other
cases, error 0003 is returned.
Setting command
LXI p1 p2 p3 p4 p5
Parameter Content
p1 = 1 to 99
L
N
1
setting
p2 = 1 to 99
L
N
2
setting
p3 = 1 to 99
L
N
3
setting
p4 = 1 to 99
L
N
4
setting
p5 = 1 to 99
L
N
5
setting
Request command
LXI ?
Response data
d1,d2,d3,d4,d5
Return value
Content
* Same as for setting command
ADP
Sub channel additional processing (corresponds to Sub channel
measurement
L
peak
/
L
tm5
)
Only valid if the unit is currently in
L
p
mode (except pause). In other
cases, error 0003 is returned.
Setting command
ADP p1
Parameter Content
p1 = 0
OFF
p1 = 1
L
peak
p1 = 2
L
tm5
Request command
ADP ?
Response data
d1
Return value
Content
* Same as for setting command
* When the setting is made via remote communication, the
sound level meter will accept the ADP command also if it is
in analyzer mode.