48
Commands
MXD
Peak hold (corresponds to Display MAX hold)
Only valid if the unit is currently in
L
p
mode (except pause). In other
cases, error 0003 is returned.
Setting command
MXD p1
Parameter Content
p1 = 0
OFF
p1 = 1
ON
* Peak hold display is performed only in analyzer mode, not in
sound level meter mode.
* When the setting is made via remote communication, the
sound level meter will accept the MXD command also if it is
in sound level meter mode.
Request command
MXD ?
Response data
d1
Return value
Content
* Same as for setting command
LNM
L
N
mode (corresponds to Measurement
L
N
mode)
Only valid if the unit is currently in
L
p
mode (except pause). In other
cases, error 0003 is returned.
Setting command
LNM p1
Parameter Content
p1 = 0
L
p
p1 = 1
L
eq, 1s
* Selecting Japanese as language is not valid (error 0003 will
be returned).
Request command
LNM ?
Response data
d1
Return value
Content
* Same as for setting command
* If Japanese is selected as language, p1 = 0 is returned (opera-
tion also corresponds to p1 = 0).