62
Commands
Calibration commands
CAL
Calibration status (corresponds to calibration mode on/off by CAL
key operation and selection of internal/acoustic calibration by
STORE key operation)
Only valid if the unit is currently in
L
p
mode (except pause) and
calibration mode. In other cases, error 0003 is returned.
Setting command
CAL p1
Parameter Content
p1 = 0
Cancel calibration mode
p1 = 1
Activate internal calibration
p1 = 2
Activate acoustic calibration
Request command
CAL ?
Response data
d1
Return value
Content
* Same as for setting command
CBM
Calibration volume (corresponds to
key operation during
calibration)
Only valid if the unit is currently in calibration mode. In other cases,
error 0003 is returned.
Setting command
CBM p1
Parameter Content
p1 = 0
Set CAL volume to minimum
p1 = 1
Set CAL volume to maximum
* If p1 = 0 is sent when CAL volume is at minimum, error 0002
is returned.
* If p1 = 1 is sent when CAL volume is at maximum, error 0002
is returned.
* To obtain the displayed value during calibration, use the DOD
command described on page 82 to 84.
* Perform calibration with the CBM command while checking the
displayed value with the DOD command.