40
Commands
Command Send Example
To set frequency weighting to “C”
(1) Start of transfer data and command
Always <
STX
>
(2) ID number (hexadecimal). The ID number range is 0 to 255. In a
command string, this is expressed as 01
H
(= ID number 1) to FF
H
(=
ID number 255).
Note
Do not use ASCII “01” here. Express “01” in 2-digit
hexadecimal notation, i.e. 01
H
.
(3) Attribute (“C” for command)
Always C
(4) Command
3 alpha characters
(5) Parameter (corresponds to p1, p2, etc. in command description section
of manual [following pages])
(6) Command end
Always <
ETX
>
(7) Always 00
H
(8) Transfer data end
Always <
CR
><
LF
>
<
STX
>
01
H
WGT
0_2
<
CR
><
LF
>
00
H
<
ETX
>
C
(1)
(2) (3)
(4)
(5)
(6)
(7)
(8)