64
Commands
PLP
Auto 1 store cycle (corresponds to Store Sampling cycle)
Only valid if the unit is currently in
L
p
mode (except pause). In other
cases, error 0003 is returned.
Setting command
PLP p1 p2
Parameter Content
p1 = 1 to 9
1 ms to 10 ms, in 1-ms steps
p1 = 10 to 1000
10 ms to 1000 ms, in 10-ms steps
p1 = 0
L
eq,1s
(use “0” to specify 1-second
L
eq
)
p2 = 0
100 ms (in sound level meter mode, only “0”
can be specifi ed)
* In analyzer mode, sampling cycle is p1.
In sound level meter mode, sampling cycle is p2.
* Between 10 and 1000, use steps of 10. For example, “25” can-
not be specifi ed.
* In sound level meter mode, the setting is fi xed to 100 ms and
cannot be changed.
Request command
PLP ?
Response data
d1,d2
Return value
Content
* Same as for setting command