59
Commands
MKP
Cursor position (corresponds to cursor positioning operation in
analyzer mode)
If the unit is displaying a menu or is in recall, calibration, or adjust-
ment mode, the command is not valid and error 0003 is returned.
In sound level meter (SLM) mode, the command is not valid and er-
ror 0003 is returned.
Setting command
MKP p1 p2
Parameter Content
p1 = 0
Not available (Do not specify this.)
p1 = 1
Not available (Do not specify this.)
p1 = 2
16 Hz band
p1 = 3
31.5 Hz band
p1 = 4
63 Hz band
p1 = 5
125 Hz band
p1 = 6
250 Hz band
p1 = 7
500 Hz band
p1 = 8
1 kHz band
p1 = 9
2 kHz band
p1 = 10
4 kHz band
p1 = 11
8 kHz band
p1 = 12
16 kHz band
p2 = 0
Bottom band
p2 = 1
Center band
p2 = 2
Top band
* In 1/3 octave analysis mode/simultaneous analysis mode, p1
specifi es the octave band unit, and p2 the 1/3 octave band unit
as bottom/center/top.
* For example, to specify the 1.25 kHz band, use p1 = 8 and p2 =2.
* Use p1 = specifi ed frequency and p2 = 1 when set the octave band.
Request command
MKP ?
Response data
d1,d2
Return value
Content
* Same as for setting command