46
Commands
MTI
Set measurement time (corresponds to Measurement Measure-
ment time)
Only valid if the unit is currently in
L
p
mode (except pause). In other
cases, error 0003 is returned.
If the time setting results in a contradictory or invalid condition, er-
ror 0002 is returned.
Setting command
MTI p1 p2
Parameter Content
p1 = 1 to 1000
Value
p2 = 0
s (Seconds)
p2 = 1
m (Minutes)
p2 = 2
h (Hours)
* The maximum allowable setting time is 24 hour when store
mode is Manual or Auto2.
Request command
MTI ?
Response data
d1,d2
Return value
Content
* Same as for setting command
* d1
specifi es the value and d2 specifi es the unit.
BER
Set back-erase (corresponds to Measurement Back-erase)
Only valid if the unit is currently in
L
p
mode (except pause). In other
cases, error 0003 is returned.
Setting command
BER p1
Parameter Content
p1 = 0
Back-erase function off
p1 = 1
5 s
Request command
BER ?
Response data
d1
Return value
Content
* Same as for setting command