68
Commands
System commands
BAT
Battery status (no corresponding operation)
This command is valid in any operation state of the unit.
Request command
BAT ?
Response data
d1,d2
Return value
Content
d1 = 1
Flashing (Empty)
d1 = 2
1 segment (Low)
d1 = 3
2 segments (Mid)
d1 = 4
3 segments (High)
d1 = 5
4 segments (Full)
d2 = 1
Unit is powered by internal batteries
d2 = 2
Unit is powered by external DC supply
* There is no setting command.
CLK
Current date/time (corresponds to System Current Date/Time)
The setting command is only valid if the unit is currently in
L
p
mode
(except pause). In other cases, error 0003 is returned.
The request command is not valid if the unit is displaying a menu or is
in recall or calibration mode. In such cases, error 0003 is returned.
Setting command
CLK p1 p2 p3 p4 p5 p6
Parameter Content
p1 = 2000 to 2063 4-digit calendar year
p2 = 1 to 12
Month
p3 = 1 to 31
Day
p4 = 0 to 24
Hours
p5 = 0 to 59
Minutes
p6 = 0 to 59
Seconds
Request command
CLK?
Response data
d1,d2,d3,d4,d5,d6
Return value
Content
* Same as for setting command