Commands
TRG
Trigger (corresponds to Measurement Trigger mode Trigger
mode setting)
Only valid if the unit is currently in
L
p
mode (except pause). In other
cases, error 0003 is returned.
Setting command
TRG p1
Parameter Content
p1 = 0
OFF
p1 = 1
Level 1
p1 = 2
Level 2
p1 = 3
Time
p1 = 4
EXT.
Request command
TRG ?
Response data
d1
Return value
Content
* Same as for setting command
LTR
Trigger level (corresponds to Measurement Trigger mode
Trigger level setting)
Only valid if the unit is currently in
L
p
mode (except pause). In other
cases, error 0003 is returned.
Setting command
LTR p1 p2
Parameter Content
p1 = 25 to 130
Trigger level for level 1 and level 2 trigger
(1-dB steps)
p2 = 0
+ slope (Use p2 = 0 in level 2 trigger )
p2 = 1
− slope
Request command
LTR ?
Response data
d1,d2
Return value
Content
* Same as for setting command
* d1 indicates trigger level, and d2 indicates trigger slope.