Protective conductor connection
The ADG must never be operated without a protective conductor connection.
In addition, a protective conductor of at least 2.5 mm² must be led to every
ultrasound converter of the ADG generator.
Potential equalisation
All components of an ultrasound system must be connected to a main
potential. They can be connected using a mounting plate such as the ADG
generator or a cable of at least 6 mm
2
Ensuring operational readiness
▪ Plug device connectors into the sockets on the generator (see figure
below)
Fig. 36:
Generator (connections)
1.
Press
2.
Two-hand start
3.
RF (converter connection)
4.
Power (mains connection)
5.
RS232 (9-pin SUB-D;
Switch data set function [
}
73]
Data string [
}
84]
)
6.
Digital input (25-pin SUB-D, control cable
sound enclosure SSK)
7.
Digital output (25-pin SUB-D, control cable
sound enclosure SSK)
Pin assignment see chapter
Pin assignment [
}
127]
.