2-73
User’s Guide for DG1000 Series
1.
Test
Press Utility
Æ
Test/Cal
Æ
Test, to perform self-testing. When the instrument is
powered, the system will perform the self-testing automatically. This limited test can
test if the instrument could work ordinarily. To perform a general test needs 15
seconds. If all the tests have been passed, you can use the generator freely. If the
general test has been passed, the screen will show “Self-Test Passed”; if not, it will
also tell you self-test failed and display the error message.
2.
Calibration
The information consist the calibration counting and the edition.
Please note that the generator has been calibrated before it leaved the factory,
When you turn on a new instrument, the default number of the calibration is 0.
You can get the information of the latest version on the company’s website:
http://www.rigol.com
.
3.
Password
This function is used to prevent unauthorized or accidental calibration to the
generator. When the generator is firstly shipped to the user, it is secured. You have to
unlock it before calibrating the instrument.
Press Utility
Æ
Test/Cal
Æ
PassWd, and enter the right password. The system will
display “The System is now unsecured”. SecOff will appear, see Figure 2-92, The
default password is “
12345
”. It is stored in the Non-Volatile Memory, and will not
change when the instrument is powered off or the Remote I/O is reset.
Figure 2-92 Operation Menu